User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: SUB FM RESIDUAL Measurement DATE: 05-Jun-94 AUTHOR: User Contributed REVISION: .1 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 30 CONFIGURATION: HP 8901A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 IEEE [@8901][TERM LF]M5 AU[I] 1.002 IEEE [@8901][MEM]/1E6 F1 M2 2.1SP L2 H1 D4 T0[I] 1.003 MATH MEM1=-50 1.004 IEEE [@8901][I] 1.005 MATH MEM1=MEM1+1 1.006 JMPZ 1.008 1.007 JMP 1.004 1.008 IEEE [@8901][D500][I] 1.009 MATH MEM1=MEM 1.010 IEEE [@8901][D500][I] 1.011 MATH MEM1=MEM1+MEM 1.012 IEEE [@8901][D500][I] 1.013 MATH MEM1=MEM1+MEM 1.014 IEEE [@8901][D500][I] 1.015 MATH MEM1=MEM1+MEM 1.016 IEEE [@8901][D500][I] 1.017 MATH MEM=(MEM1+MEM)/5